If using the handheld local controller, move the needle down until the needle tip is approx. 2–4 mm above the seat. This will allow access to the needle assembly to remove the sample loop. Disconnect the sample loop from the needle, and, if needed, also from the top of the Analytical Head. (See Figure 4) Figure 4.